| Model Number |
IPN048992 |
| Medical Device Problem Codes |
Fracture (1260); Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).
|
| |
|
Event or Problem Description
|
|
Complaint reports that: "blue plastic tube connected to the main body is loose.Not connected properly.".
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#: (b)(4).Complaint verification testing could not be performed as it was reported that the sample is not available for return.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|
Event or Problem Description
|
|
Complaint reports that: "blue plastic tube connected to the main body is loose.Not connected properly." the issue was detected prior to use on a patient.
|
| |
|
Search Alerts/Recalls
|