MAUDE Adverse Event Report: ZIMMER BIOMET, INC. 6.0 MM DIAMETER CANCELLOUS SCREW FIXED ANGLE 3.5 MM HEX HEAD; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. 6.0 MM DIAMETER CANCELLOUS SCREW FIXED ANGLE 3.5 MM HEX HEAD; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 47248405060
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Hematoma (1884); Sepsis (2067); Impaired Healing (2378)
Event Type  Injury  
Event Description
It was reported via a clinical study that a patient underwent a bone fixation procedure of the right femur.On an unknown time frame post-implantation, the patient developed a traumatic right femoral hematoma with a wound healing disorder.Blood cultures tested positive for e.Coli in sepsis and a smear also tested positive for e.Coli.Subsequently, the patient underwent medical intervention including a jet lavage procedure to clear the hematoma and wound.Due diligence is complete as multiple attempts have been made; however no further information is available and the product remains implanted.
 
Manufacturer Narrative
(b)(4).Event occurred in 2015 though exact date is unknown.Implantation occurred in 2015 though exact date is unknown.Concomitant medical products: retrograde femoral nail - purple 10 mm diameter 36 cm length: catalog#47249436010, lot#61295569; 6.0 mm diameter cancellous screw - black fixed angle 3.5 mm hex head: catalog#47248405060, lot#61143263; 6.0 mm diameter cancellous screw - black fixed angle 3.5 mm hex head: catalog#47248405560, lot#61222158; 6.0 mm diameter cancellous screw - black fixed angle 3.5 mm hex head: catalog#47248406560, lot#61374598; 5.0 mm diameter cortical screw - red fixed angle 3.5 mm hex head: catalog#47248403250, lot#61812273; 5.0 mm diameter cortical screw - red fixed angle 3.5 mm hex head: catalog#47248403750, lot#61711476.Foreign source: germany.Multiple mdr reports have been filed for this event.Please see associated reports: 0001822565-2022-03012; 0001822565-2022-03013; 0001822565-2022-03014; 0001822565-2022-03015; 0001822565-2022-03017; 0001822565-2022-03018.Customer has indicated that the product will not be returned to zimmer biomet for evaluation, as it remains implanted.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.D10: medical product: retrograde femoral nail - purple 10 mm diameter 36 cm length: catalog#: 47249436010, lot#: 61295569; 6.0 mm diameter cancellous screw - black fixed angle 3.5 mm hex head: catalog#: 47248405060, lot#: 61143263; 6.0 mm diameter cancellous screw - black fixed angle 3.5 mm hex head: catalog#: 47248405560, lot#: 61222158; 6.0 mm diameter cancellous screw - black fixed angle 3.5 mm hex head: catalog#: 47248406560, lot#: 61374592; 5.0 mm diameter cortical screw - red fixed angle 3.5 mm hex head: catalog#: 47248403250, lot#: 61812273; 5.0 mm diameter cortical screw - red fixed angle 3.5 mm hex head: catalog#: 47248403750, lot#: 61711476.The reported infection/sepsis that developed was identified as positive for e.Coli bacteria, which was obtained via culture of the hematoma.A hematoma is a mass of clotted blood that forms in a tissue, organ, or body space.A hematoma can be associated with pain, swelling, ecchymosis, serosanguinous or bloody drainage after a recent surgical procedure or trauma and can predispose the patient to infection.The development of a postoperative hematoma can be correlated with the surgical procedure and perioperative anticoagulation therapy prescribed to prevent thrombus formation.Most hematomas resolve on their own, without surgical intervention, while some do not.Larger hematomas may need to be surgically evacuated in order to resolve.As timeframes of onset differ due to individual contributing factors, a specific timeframe of expected occurrence cannot be established.Investigation results concluded that the reported event was due to non-device related factors.Device history record was reviewed and no discrepancies related to the reported event were found.Sterile certifications were also reviewed and found to be conforming with no applicable deviations.Devices were verified to have gone through acceptable sterilization process following iso/aami/astm & eu published guidelines.There are multiple factors that may contribute to an infection that are outside the control of zimmer biomet, such as external factors, i.E.Hospital/surgical environment, provider related risk factors, and/or patient comorbidities/risk factors.As there are no indications of a product or process issues identified affecting implant safety or effectiveness, implanted products are not identified as the source or contributing to the reported infection.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information at time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
6.0 MM DIAMETER CANCELLOUS SCREW FIXED ANGLE 3.5 MM HEX HEAD
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15650059
MDR Text Key302174942
Report Number0001822565-2022-03016
Device Sequence Number1
Product Code HSB
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K083497
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Study,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/21/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/31/2019
Device Catalogue Number47248405060
Device Lot Number61236378
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/03/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/20/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SEE H10 NARRATIVE.; SEE H10 NARRATIVE.
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-