Model Number CNA0T0
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
A nurse reported with a description of intraocular lens (iol) not cooperating during the procedure.There was a patient contact noted.There was no patient harm reported.The procedure was completed with another iol during initial procedure.Additional information has been requested.
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
 
Additional Manufacturer Narrative
The product was returned for analysis and the reported complaint could not be observed.Iol (intraocular lens) was returned outside of the device.No damage was observed to the lens or to the device.The device was returned in the blister tray in the carton.The lock-out assembly has been removed.The plunger is oriented correctly.The plunger has been fully advanced outside the tip of the device.Viscoelastic is dried in the device.No damage or abnormalities observed.Iol was returned loose in the carton.Solution is dried on the iol.No damage observed.The product investigation could not identify the root cause for the reported complaint "lens not cooperating ".No damage was observed to the returned product.The reported complaint is lacking in relevant information such as the type of defect observed to complete a thorough investigation.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON IOL WITH THE AUTONOME DELIVERY SYSTEM
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15650186
Report Number9612169-2022-00549
Device Sequence Number16751177
Product Code HQL
UDI-Device Identifier00380652396462
UDI-Public00380652396462
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 01/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2024
Device Model NumberCNA0T0
Device Lot Number25262950
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/28/2022
Supplement Date Received by Manufacturer12/08/2022
Initial Report FDA Received Date10/21/2022
Supplement Report FDA Received Date01/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/02/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DUOVISC OPHTHALMIC VISCOSURGICAL DEVICE
Patient SexUnknown
-
-