Additional information provided in d.9., h.3., h.6., and h.10.The device with the lens was returned inside the opened blister tray inside the opened carton.The lens stop and plunger lock were removed.The plunger was oriented correctly.Viscoelastic was observed in the device.The plunger was retracted into the loading area.The lens was located in the nozzle area.The leading haptic was in a straight position.This position is acceptable per the ifu (instructions for use).The trailing haptic was positioned on the anterior optic surface.The distal area of the trailing haptic was broken.The broken tip was oriented toward the nozzle area which may suggest the distal tip area had looped over or across the plunger tip during delivery attempt.The trailing optic edge was torn by the advancing plunger.The nozzle was removed and cleaned for further evaluation.The lens was removed during cleaning.Top coat dye stain testing was conducted with acceptable results.Viscoelastic information was not provided.It is unknown if a qualified viscoelastic was used.The root cause cannot be determined for the reported complaint of straight leading haptic and lens damaged in delivery system.The optic was broken at the trailing edge.The trailing haptic was misfolded and broken onto the optic.The root cause for the misfolded trailing haptic cannot be determined.Haptic folding may be influenced by an intermittent plunger rate or if the operating room temperature is too high (> 23°c / 73° f).If the operating room temperature is too high lens folding consistency is negatively affected as the lens more adherent.This may inhibit lens advancement or contribute to incorrect haptic folding.The broken trailing optic edge and broken trailing haptic suggest that a plunger override may have occurred during the delivery attempt.It is unknown if a qualified viscoelastic was used.The ifu instructs: during device preparation and implantation of the company iol with the company preloaded delivery system, an company qualified ophthalmic viscoelastic device (ovd) should be used.The use of an unqualified ovd may cause damage to the lens and potential complications during the device preparation and implantation steps.Plunger override may occur: due to rapid advancement faster than the ifu recommend rate.Due to the use of a non-qualified viscoelastic.Material properties of non-qualified.Ovds may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.If inadequate viscoelastic is placed in the device this will cause inadequate coverage of the lens fold path which may cause the lens to advance incorrectly or become ¿stuck¿ in the device allowing the plunger to override the lens.If the device is overfilled with ovd, this can prevent the trailing haptic from being placed properly or move the lens out of position resulting in misfolding.If the operating room temperature is too high (> 23°c / 73° f) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement.A straight leading haptic was also observed.Straight leading haptics are not a product malfunction.This is an acceptable position per the diagrams provided in the ifu.It is unknown if a qualified viscoelastic was used.The ifu instructs: during device preparation and implantation of the acrysof¿ iq iol with the ultrasert¿ preloaded delivery system, an alcon qualified ophthalmic viscoelastic device (ovd) should be used.The use of an unqualified ovd may cause damage to the lens and potential complications during the device preparation and implantation steps.Top coat dye stain testing was conducted with acceptable results.A straight leading haptic may occur: due to the normal folding variations as indicated the ifu diagrams.If the initial plunger advancement is faster than the recommended target, the leading haptic may be forced past the internal folding feature.If there is excessive delay between the device preparation and subsequent delivery the leading haptic may start to unfold.Due to the use of a non-qualified viscoelastic.Material properties of non-qualified ovds may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.If the operating room temperature is too high (> 23°c / 73° f) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement or contribute to incorrect haptic folding.Any of the above listed causes alone, or in combination, may create the reported event.The manufacturer internal reference number is: (b)(4).
|