MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. ENDOSCOPE POSITION DETECTING UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. ENDOSCOPE POSITION DETECTING UNIT Back to Search Results
Model Number UPD-3
Medical Device Problem Code Loose or Intermittent Connection (1371)
Health Effect - Clinical Code Pain (1994)
Date of Event 09/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device referenced in this report was not returned to olympus for evaluation.The definitive cause of the user's experience cannot be determined at this time.The investigation is ongoing.This report will be updated upon completion of the investigation or upon receipt of additional relevant information.
 
Event or Problem Description
The customer reports during an unspecified procedure using an endoscopic position detecting device with a video system center, the user was unable to capture an image.There was a 45-minute delay in the procedure which left the patient in pain.The user had to capture the image manually.No additional consequences to the patient were reported as a result of this occurrence.The equipment was inspected onsite by medical engineering.Medical engineering confirmed that the scope guide had a loose connection which was resolved by the hospital¿s staff on site.The video system center had issues with the settings which has now been rectified using tech support.With that said medical engineering believe that the issue did not lie with the equipment.The equipment is not being returned to olympus for evaluation since the issues were corrected onsite.Case with patient identifier (b)(6) reports the video system center used in the procedure.Case with patient identifier (b)(6) reports the endoscopic position detecting device used in the procedure.
 
Additional Manufacturer Narrative
This report is being updated to report corrected information.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been 2 years since the subject device was manufactured.Based on the results of the investigation, since the device was not returned for an evaluation, the root cause of the reported event could not be determined.However, it is likely that the following occurred: patient was left in pain due to the image issue, an issue with the processor and scope guide, and/or there was a clinically relevant delay occurred 45 minutes.This supplemental report includes a correction to d4.Udi number inadvertently not included on the initial medwatch.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE POSITION DETECTING UNIT
Common Device Name
ENDOSCOPE POSITION DETECTING UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer Contact
masaharu hirose
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8-061
JA   961-8061
426422891
MDR Report Key15650453
Report Number3002808148-2022-03418
Device Sequence Number2428437
Product Code PGU
UDI-Device Identifier04953170331145
UDI-Public04953170331145
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K134026
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberUPD-3
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/26/2022
Supplement Date Received by Manufacturer02/02/2023
02/02/2023
Initial Report FDA Received Date10/21/2022
Supplement Report FDA Received Date10/21/2022
02/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/16/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-