Catalog Number C45101340-NLJ
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/01/2018
Event Type  malfunction  
Manufacturer Narrative
Udi and (510k) are unknown; no additional information was provided.This mdr was generated under protocol (b)(4)., as a result of warning letter cms# (b)(4).A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.The sample was received opened in its original packaging.Visual inspection: the sample was visually inspected at a distance of 12 inches under normal lighting; found a hole in the bag tip.The most probable root cause of the issue was deemed to be supplier-related; bag was sent by the supplier in defective conditions and production personnel did not detect this issue during the assembly process.Action(s) taken to mitigate the reported issue(s): supplier was notified, a qalert was generated to ensure that operators verify that breathing bags are free of holes and damaged, and production personnel was notified by quality intern as awareness of the defect reported by the customer.
 
Event Description
It was reported that, immediately after opening the package, the customer noticed the end part of the anesthesia bag was torn.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX GENERAL ANESTHESIA CIRCUITS
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15650772
MDR Text Key307093978
Report Number3012307300-2022-25698
Device Sequence Number1
Product Code BTO
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/01/2021
Device Catalogue NumberC45101340-NLJ
Device Lot Number3598040
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/17/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/03/2022
Initial Date FDA Received10/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/02/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-