• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STERIS CORPORATION - MONTGOMERY HARMONYAIR A-SERIES SURGICAL LIGHTING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STERIS CORPORATION - MONTGOMERY HARMONYAIR A-SERIES SURGICAL LIGHTING SYSTEM Back to Search Results
Model Number A-SERIES LIGHTING SYSTEM
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/23/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A steris service technician arrived onsite to inspect the lighting system and found that the yoke assembly had come loose from the female adapter resulting in the reported event.The technician replaced the dual monitor yoke assembly, tested the unit, confirmed to be operating according to specifications, and returned it to service.The damaged parts were returned to montgomery for engineering to review.A follow-up mdr will be submitted when additional information becomes available.
 
Event or Problem Description
The user facility reported that their dual monitor yoke assembly detached from the suspension arm of their harmonyair a-series surgical lighting system and was hanging by the wiring within the light.There was no one in the room at the time of the reported event.No report of injury or procedural delay.
 
Additional Manufacturer Narrative
The damaged yoke assembly was returned for evaluation.A root cause of the reported event could not be determined.Steris will continue to monitor for similar events.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONYAIR A-SERIES SURGICAL LIGHTING SYSTEM
Common Device Name
LIGHTING SYSTEM
Manufacturer (Section D)
STERIS CORPORATION - MONTGOMERY
2720 gunter park drive east
montgomery AL 36109
Manufacturer (Section G)
STERIS CORPORATION - MONTGOMERY
2720 gunter park drive east
montgomery AL 36109
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key15651222
Report Number1043572-2022-00070
Device Sequence Number6241139
Product Code FSY
UDI-Device Identifier00724995187637
UDI-Public00724995187637
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA-SERIES LIGHTING SYSTEM
Device Catalogue NumberLAH11IN02
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/23/2022
Supplement Date Received by Manufacturer09/23/2022
Initial Report FDA Received Date10/21/2022
Supplement Report FDA Received Date01/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/06/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-