• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL GUIDE WIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL GUIDE WIRE Back to Search Results
Model Number 1009660J
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code Vascular Dissection (3160)
Date of Event 09/29/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the procedure was to treat a lesion in the right coronary artery (rca).The balance middleweight (bmw) universal guide wire distal tip (9.5cm) separated in the anatomy.There was no resistance during advancement and removal.After extensive work the distal tip was retrieved with a snare device.Nothing angiographically detectable, no evidence of echocardiogram (ecg) changes or changes of symptoms confirm nothing was left in the patient.A dissection occurred during snaring and retrieval of the distal end of the guide wire.There was no adverse patient sequela; however, there was a clinically significant delay reported.No additional information was provided.
 
Additional Manufacturer Narrative
The device is expected to be returned for evaluation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the procedure was to treat a lesion in the right coronary artery (rca).The balance middleweight (bmw) universal guide wire distal tip (9.5cm) separated in the anatomy.There was no resistance during advancement and removal.After extensive work the distal tip was retrieved with a snare device.Nothing angiographically detectable, no evidence of echocardiogram (ecg) changes or changes of symptoms confirm nothing was left in the patient.A dissection occurred during snaring and retrieval of the distal end of the guide wire.There was no adverse patient sequela; however, there was a clinically significant delay reported.Subsequent to the initially filed report, the following information was provided.The dissection was treated with an additional stent in the distal rca.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record and similar incident review of the reported lot could not be conducted because the lot number was not provided.As there was no damage noted to the device during the inspection prior to use, the investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that during use inadvertent mishandling and/or interaction with the anatomy/other devices resulted in the reported tip material separation.The reported treatment possibly caused/contributed to the reported patient effect; however a conclusive cause for the reported patient effect, and the relationship to the product, if any, cannot be determined.The treatment appears to be related to the operational context of the procedure as the distal tip was retrieved with a snare device.The reported patient effect of dissection is listed in the hi-torque guide wires for ptca, pta, and stents instructions for use as a known patient effect of coronary procedures.There is no indication of a product quality issue with respect to manufacture, design or labeling.D10, h3 - it was initially reported that the device would be returned for analysis.Subsequent information revealed that the device was discarded and is not available for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL GUIDE WIRE
Common Device Name
GUIDE WIRE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005737652
road no.2 km 58.0 cruce dávila
barceloneta PR 00617
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15651344
Report Number2024168-2022-10901
Device Sequence Number6094385
Product Code DQX
UDI-Device Identifier08717648065866
UDI-Public08717648065866
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K101011
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1009660J
Device Catalogue Number1009660J
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/03/2022
Supplement Date Received by Manufacturer01/12/2023
Initial Report FDA Received Date10/21/2022
Supplement Report FDA Received Date01/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-