• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MELSUNGEN AG INTROCAN SAFETY®; CATHETER,INTRAVASCULAR,THE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MELSUNGEN AG INTROCAN SAFETY®; CATHETER,INTRAVASCULAR,THE Back to Search Results
Model Number 4252519-02
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event or Problem Description
As reported by the user facility: detailed inquiry description: the nurse noticed that there is a debris or a plastic like speck on the side of a brand new iv catheter when he opened the package and took the cap off the needle.No injury reported.
 
Additional Manufacturer Narrative
Event 1: this report has been identified as b.Braun medical internal report number (b)(4).One (1) unused sample in opened packaging and photo were provided for evaluation.Visual inspection of the photograph notes a white fiber like contaminant found on the capillary surface.The actual origin of contaminant was unable to be identified from this picture.Through visual inspection of the sample, it was observed that no similar white contaminant was on the returned sample and/or the packaging.Further analysis on the origin of the particle contaminate is not possible.A review of manufacturing records was performed and indicated that there were no quality issues during the manufacturing of this lot related to the reported issue.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTROCAN SAFETY®
Common Device Name
CATHETER,INTRAVASCULAR,THE
Manufacturer (Section D)
B. BRAUN MELSUNGEN AG
carl-braun strabe 1
melsungen, 34212
GM  34212
Manufacturer (Section G)
B. BRAUN MELSUNGEN AG
carl-braun strabe 1
melsungen, 34212
GM   34212
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key15652898
Report Number9610825-2022-00426
Device Sequence Number14961435
Product Code FOZ
UDI-Device Identifier04046963165949
UDI-Public(01)04046963165949
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K020785
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number4252519-02
Device Catalogue Number4252519-02
Device Lot Number22E15G8361
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/14/2022
Supplement Date Received by Manufacturer10/14/2022
Initial Report FDA Received Date10/21/2022
Supplement Report FDA Received Date02/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-