| Catalog Number |
24355 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A hospital alleged multiple 3m¿ ranger¿ high flow warming sets (model 24355) leaked during use.Location of leak was identified to be the bubble trap and at the leur connections.No patient or staff injury was alleged.No medical interventions were required.
|
| |
|
Additional Manufacturer Narrative
|
|
Patient specifics are unknown.Date of event is unknown.Product has not been returned to 3m¿ for analysis.3m¿ is unable to verify the leak and root cause without the sample.3m¿ will continue to monitor.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: 3m¿ has investigated and identified a manufacturing issue with the auto-venting bubble trap that introduces the risk of blood or fluid leaks while priming the sets and/or during fluid administration.There is no alleged injury to the patient or caregiver for this reported incident.3m¿ has initiated a field safety corrective action for lots associated with this issue.3m¿ has implemented a change to address the identified manufacturing issue and will continue to monitor.
|
| |
|
Search Alerts/Recalls
|