• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILLIAM COOK EUROPE FROVA INTUBATING INTRODUCER; Tube, tracheal (w/wo connector)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM COOK EUROPE FROVA INTUBATING INTRODUCER; Tube, tracheal (w/wo connector) Back to Search Results
Model Number G48303
Medical Device Problem Codes Component Incompatible (1108); Migration (4003)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Serious Injury
Event or Problem Description
According to initial reporter: piece of plastic sheared off when tube placed over, ending up in patient's airway.
 
Additional Manufacturer Narrative
Btr tube, tracheal (w/wo connector).
 
Event or Problem Description
Additional information provided states: the small piece of catheter was retrieved and will not affect patient outcomes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FROVA INTUBATING INTRODUCER
Common Device Name
Tube, tracheal (w/wo connector)
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
MDR Report Key15654212
Report Number3005580113-2022-00110
Device Sequence Number6387302
Product Code BTR
UDI-Device Identifier00827002483033
UDI-Public(01)00827002483033(17)221010(10)E3908347
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/08/2022
Report Date (Section F) 10/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/10/2022
Device Model NumberG48303
Device Catalogue NumberC-CAE-14.0-70-FIC
Device Lot NumberE3908347
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date10/28/2022
Device Age3 YR
Event Location Hospital
Date Report to Manufacturer10/28/2022
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date10/21/2022
Supplement Report FDA Received Date11/08/2022
Date Device Manufactured10/10/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-