Model Number 670175
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code Decreased Respiratory Rate (2485)
Date of Event 09/20/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A product sample was received and is awaiting evaluation, investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event or Problem Description
It was reported that the trach tube had cuff leakage that was discovered around 72 hours after insertion.The issue was resolved with a trach tube change.
 
Additional Manufacturer Narrative
H10: device evaluation: one sample was returned for investigation.The sample was received in used condition without its original packaging open and inside in a plastic bag.The sample was visually inspected under normal conditions of illumination to detect any damage on the cuff, airline, or pilot balloon.The sample received did not present any damaged, kink, cut that could cause the failure mode reported.The sample was then tested for leak by introducing water in the product.A rip was found on the cuff; therefore, a leak was found; complaint was confirmed.By the physical evaluation of the unit, the complaint was confirmed.No other analysis was performed.The cause of the failure mode was traced to manufacturing.Capa was opened to investigate and address the root cause and corrective actions relating to inflation, leakage, and symmetrical issues.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA TRACHEOSTOMY TUBE, ADULT
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
NULL
5700 west 23rd ave
gary IN 46406
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15654735
Report Number3012307300-2022-25819
Device Sequence Number2530595
Product Code JOH
UDI-Device Identifier15021312006063
UDI-Public15021312006063
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K083641
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number670175
Device Catalogue Number670175
Device Lot Number4229360
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/22/2022
Supplement Date Received by Manufacturer01/16/2023
Initial Report FDA Received Date10/21/2022
Supplement Report FDA Received Date05/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/14/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
Patient EthnicityNon Hispanic
Patient RaceAsian
-
-