• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CHRISTOPH MIETHKE GMBH & CO KG PROGAV 2.0 WITH SA20 AND DISTAL CATHETER; HYDROCEPHALUS MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHRISTOPH MIETHKE GMBH & CO KG PROGAV 2.0 WITH SA20 AND DISTAL CATHETER; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FX420T
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 09/22/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
It was reported that a progav 2.0 shunt system (part # fx420t) was implanted during a procedure performed on (b)(6) 2022.According to the complainant, the shunt system was believed to be blocked.The patient underwent a revision procedure performed on (b)(6) 2022.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Same event as medwatch# 3004721439-2022-00319.
 
Additional Manufacturer Narrative
Investigation: visual inspection: the following observations were made during the visual inspection: -discoloured areas at the catheter -blood and tissue residues permeability test: the test showed that the progav 2.0 shunt system is non-permeable.The ventricular catheter is permeable.To determine the location of the occlusion, the shunt system was dismantled and each element was tested individually for permeability.The test showed that the progav 2.0 is non-permeable, whereas the shuntassistant is permeable.Computer control test: due to the non-permeability of the progav 2.0, the computer-controlled test is not applicable.The shuntassistant was tested according to standard procedure in the vertical position of the valve.At a fixed opening pressure of 20 cmh2o in the vertical position, a pressure of 20 cmh2o +4/-2 cmh2o is expected.The results indicated that at a reference flow of 20 ml/h in the vertical position, the shuntassistant showed an opening pressure of 7,12 cmh2o.This is not within the specified tolerance of 20 cmh2o +4/-2 cmh2o.According to our results, we can detect an accelerated outflow.Adjustability test: the progav 2.0 was found to be adjustable to all pressure settings.The shuntassistant is a fixed pressure valve, therefore the adjustability test is inapplicable.Braking force and brake function test: the braking force of the progav 2.0 was within the specified tolerance and the brake function operated as expected.Internal inspection of product: in order to verify whether the investigated shunt system was compromised by the known risks of hydrocephalus therapy, e.G.By a build-up of natural substances (protein, blood, or tissue particles) in the cerebrospinal fluid, the valves are finally opened with a special tool.After dismantling of the valves, bloody deposits were found in both valves.Additionally, a material residue was found in the inlet of the shuntassistant.Results: based on our investigation results, we confirm the blockage in the shunt system.Detected deposits could be named as the cause of the blockage.Endogenous substances in the csf can temporarily impair the function of the valve and is described as concomitant symptoms in hydrocephalus therapy.Concluding on the determined accelerated outflow at the computer test bench, we can prove that the determined material residue in the shuntassistant was not the cause for the blockage in the shunt system.From our point of view, no further regulatory actions are required.
 
Event or Problem Description
Same event as medwatch# 3004721439-2022-00319.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV 2.0 WITH SA20 AND DISTAL CATHETER
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key15654913
Report Number3004721439-2022-00320
Device Sequence Number2356745
Product Code JXG
UDI-Device Identifier04041906136628
UDI-Public4041906136628
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K161853
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFX420T
Device Catalogue NumberFX420T
Device Lot Number20057327
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/23/2022
Supplement Date Received by Manufacturer10/12/2022
Initial Report FDA Received Date10/21/2022
Supplement Report FDA Received Date11/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/28/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age10 YR
Patient SexUnknown
-
-