Product complaint # (b)(4).Additional information was requested, and the following was obtained: please clarify if the needle had come loose from the thread.Occurred intra-op, or pre-op? it come loose pre-op lot number? scbbktw0.H3 investigation summary: the product was returned for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that one opened sample that pertains to the product code vcp393h was returned for analysis after investigation of the sample short insertion was observed in the strand.The visual inspection concluded that the combination of the needle/suture was found detached since the insertion end of the suture did not meet the requirement, this defect is caused because the suture was not correctly inserted inside of the needle barrel hole resulting in a pull-out defect.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.This is a combination product, and the event has been reviewed for both the suture and the triclosan.Trade name - irgacare®.Active ingredient(s) ¿ triclosan.Dosage form ¿ suture/solid/parenteral.Strength ¿ = 275 ¿g/m.This report is being submitted pursuant to the provisions of 21 cfr part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|