| Catalog Number |
MC1321P |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that an roi-c plate jammed in the cage's slot.The surgeon removed both pieces and tried again outside the patient and was able to get them to fit together with force.Neither item was re-used and the surgery continued with a delay of 20 minutes, but without impact to the patient.This is report two of two for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference report 3004788213-2022-00085.
|
| |
|
Additional Manufacturer Narrative
|
|
Inspection: the products were not returned and no photos were provided, so an evaluation is unable to be performed.Dhr review: the dhrs were reviewed.There are no indications of manufacturing issues which would have contributed to this event and the devices were likely conforming when they left zimvie¿s control.Potential root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to off-axis forces applied during use.Device usage: this device is used for treatment.A follow-up report will be sent if additional information is obtained that adds value to the relevant content of this report.
|
| |
|
Event or Problem Description
|
|
It was reported that an roi-c plate jammed in the cage's slot.The surgeon removed both pieces and tried again outside the patient and was able to get them to fit together with force.Neither item was re-used and the surgery continued with a delay of 20 minutes, but without impact to the patient.This is report two of two for this event.
|
| |
|