• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET SPINE INC. ZYSTON STRAIGHT INLINE INSERTER; ZYSTON STRAIGHT SPACER SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE INC. ZYSTON STRAIGHT INLINE INSERTER; ZYSTON STRAIGHT SPACER SYSTEM Back to Search Results
Catalog Number 14-534900
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/28/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.Reference report 3012447612-2022-00247.
 
Event or Problem Description
It was reported that the tip of a vital t27 final driver broke off in a set screw while attempting to remove a previously-placed construct.Then a tab on the tip of a zyston straight inline inserter broke off while it was being connected to a previously-placed cage.Both fragments were recovered from the patient and alternative instruments were used to complete the procedure without patient impacts.This is report one of two for this event.
 
Additional Manufacturer Narrative
Device evaluation visual inspection revealed that one of the prongs at the tip of the inserter was fractured off.Potential cause root cause was unable to be determined.This event could possibly be attributed to wear through use over time or multiple sterilization cycles.Dhr review per dhr review, the parts were likely conforming when they left zimvie control.Device use this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
 
Event or Problem Description
It was reported that the tip of a vital t27 final driver broke off in a set screw while attempting to remove a previously-placed construct.Then a tab on the tip of a zyston straight inline inserter broke off while it was being connected to a previously-placed cage.Both fragments were recovered from the patient and alternative instruments were used to complete the procedure without patient impacts.This is report one of two for this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZYSTON STRAIGHT INLINE INSERTER
Common Device Name
ZYSTON STRAIGHT SPACER SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
kim martinez
10225 westmoor dr.
na
westminster, CO 80021
3035144809
MDR Report Key15654984
Report Number3012447612-2022-00246
Device Sequence Number2428511
Product Code MAX
UDI-Device Identifier00888480472100
UDI-Public(01)00888480472100(10)B341040
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K112014
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number14-534900
Device Lot NumberB341040
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/29/2022
Supplement Date Received by Manufacturer11/27/2022
Initial Report FDA Received Date10/21/2022
Supplement Report FDA Received Date11/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/23/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexFemale
-
-