CHRISTOPH MIETHKE GMBH & CO KG CONTROLRESERVOIR SET W/DISTAL CATHETER; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FV049T
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 09/22/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
It was reported that a control reservoir (partt # fv049t) was implanted during a procedure performed on (b)(6) 2022.According to the complainant, the reservoir was believed to be blocked.The patient underwent a revision procedure performed on (b)(6) 2022.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Same event as medwatch# 3004721439-2022-00320.
 
Event or Problem Description
Same event as medwatch# 3004721439-2022-00320.
 
Additional Manufacturer Narrative
Additional information - d4 (serial #).Investigation: visual inspection: the following observations were made during the visual inspection: -bloody deposits in the control reservoir.Permeability tes: the test showed that the control reservoir is permeable.Results: based on our investigation results, we can determine deposits.The deposits had no affect to the technical properties at the time of the investigation.The cause of the aforementioned functional impairment is not known to us at the time of the examination.Proteins in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.We can exclude a defect at the time of release.The control reservoir met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.From our point of view, no further regulatory actions are required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTROLRESERVOIR SET W/DISTAL CATHETER
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key15654987
Report Number3004721439-2022-00319
Device Sequence Number2530602
Product Code JXG
UDI-Device Identifier04041906131487
UDI-Public4041906131487
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K110206
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFV049T
Device Catalogue NumberFV049T
Device Lot Number20057568
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/23/2022
Supplement Date Received by Manufacturer10/12/2022
Initial Report FDA Received Date10/21/2022
Supplement Report FDA Received Date11/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/14/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age10 YR
Patient SexUnknown
-
-