| Model Number |
21-7605-24 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during the use of the product, leakage of medical fluid from the connector was observed.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: device evaluation: two pictures were received and reviewed.Picture one shows an extension outside of its original packaging.Picture two shows a union between female luer and connector with a leak.One sample was returned for evaluation of the failure mode; the unit returned was received inside of a plastic bag with their original packaging open and decontaminated.The sample was tested using a syringe.Colored water was passed through the extension set to detect any leak; during testing, a leakage was found in the sample union between the connector and the extension tube.The complaint was confirmed.Based on the analysis conducted in the sample provided the issue was due to a lack of solvent.The failure condition was possibly due to the assembly process not being followed correctly.Failure was traced to manufacturing.For corrective action an awareness/customer complaint notification was performed to production personnel to explain the importance of adherence to manufacturing procedure a device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
|
| |
|
Search Alerts/Recalls
|