Model Number 21-7605-24
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the use of the product, leakage of medical fluid from the connector was observed.No patient injury was reported.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Additional Manufacturer Narrative
H10: device evaluation: two pictures were received and reviewed.Picture one shows an extension outside of its original packaging.Picture two shows a union between female luer and connector with a leak.One sample was returned for evaluation of the failure mode; the unit returned was received inside of a plastic bag with their original packaging open and decontaminated.The sample was tested using a syringe.Colored water was passed through the extension set to detect any leak; during testing, a leakage was found in the sample union between the connector and the extension tube.The complaint was confirmed.Based on the analysis conducted in the sample provided the issue was due to a lack of solvent.The failure condition was possibly due to the assembly process not being followed correctly.Failure was traced to manufacturing.For corrective action an awareness/customer complaint notification was performed to production personnel to explain the importance of adherence to manufacturing procedure a device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD EXTENSION SETS
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
NULL
6000 nathan lane north
minneapolis MN 55442
Manufacturer Contact
jim vegel
6000 nathan lane north
7-1-1 akasaka,
minneapolis, MN 55442
MDR Report Key15655091
Report Number3012307300-2022-25834
Device Sequence Number5965883
Product Code FPA
UDI-Device Identifier10610586044113
UDI-Public10610586044113
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K162219
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/07/2024
Device Model Number21-7605-24
Device Catalogue Number21-7605-24
Device Lot Number4279903
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/20/2022
Supplement Date Received by Manufacturer02/27/2023
Initial Report FDA Received Date10/21/2022
Supplement Report FDA Received Date06/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/21/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-