• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL AMPLATZER MULTI-FENESTRATED SEPTAL OCCLUDER - CRIBRIFORM; TRANSCATHETER SEPTAL OCCLUDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL AMPLATZER MULTI-FENESTRATED SEPTAL OCCLUDER - CRIBRIFORM; TRANSCATHETER SEPTAL OCCLUDER Back to Search Results
Model Number MF SEPTAL OCCLUDER CRIBRIFORM
Medical Device Problem Code Migration or Expulsion of Device (1395)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/23/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on (b)(6) 2022, a 35mm amplatzer multi- fenestrated septal occluder cribriform was implanted into a patient.Sizing of the device was determined by transesophageal echocardiography.During the deployment, the patient reflexively coughed, causing part of the right disc to cross the patent foramen ovale tunnel.It was attempted to move the device back into position using a lasso, but the device embolized into the left ventricle.The patient was brought to surgery and the device was removed, and the patent foramen ovale was sutured.The patient is reported to be stable and recovering.The patient remained hemodynamically stable throughout the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Additional Manufacturer Narrative
An event of device embolism was reported.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.One video which showed the patient's defect prior to the attempted device implant was received.Information from the field indicated that patient coughed during the release of the occluder, and then device migrated, which could have contributed to the reported event.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER MULTI-FENESTRATED SEPTAL OCCLUDER - CRIBRIFORM
Common Device Name
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL REG# 2135147
5050 nathan ln n
plymouth MN 55442
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key15655182
Report Number2135147-2022-01731
Device Sequence Number17651354
Product Code MLV
UDI-Device Identifier00811806010311
UDI-Public00811806010311
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
P000039
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2026
Device Model NumberMF SEPTAL OCCLUDER CRIBRIFORM
Device Catalogue Number9-ASD-MF-035
Device Lot Number8183460
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/26/2022
Supplement Date Received by Manufacturer12/19/2022
Initial Report FDA Received Date10/21/2022
Supplement Report FDA Received Date01/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age48 YR
Patient SexMale
-
-