Model Number SC-2218-70E
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/30/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the lead was fraying during a trial procedure.
 
Event or Problem Description
It was reported that the lead was fraying during a trial procedure.
 
Additional Manufacturer Narrative
Sc-2218-70e (sn: (b)(6)).The returned lead was analyzed, and visual inspection found that the lead was frayed approximately 8 cm from distal end.No cables are exposed at the damaged portion of the lead.With all the available information, boston scientific concludes the reported event was confirmed.This kind of anomaly is consistent with the damages done to a lead when the orientation of the insertion needles bevel is facing down, or the angle of the insertion needle is greater than forty-five degrees.The probable cause is due to unintended use error which caused or contributed to event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINEAR ST
Common Device Name
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
erik sherburne
25155 rye canyon loop
valencia, CA 91355
7632920920
MDR Report Key15655611
Report Number3006630150-2022-05709
Device Sequence Number11369733
Product Code LGW
UDI-Device Identifier08714729820741
UDI-Public08714729820741
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P030017
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/17/2024
Device Model NumberSC-2218-70E
Device Catalogue NumberSC-2218-70E
Device Lot Number7086956
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/30/2022
Supplement Date Received by Manufacturer10/27/2022
Initial Report FDA Received Date10/21/2022
Supplement Report FDA Received Date11/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age51 YR
Patient SexFemale
-
-