ST. JUDE MEDICAL, INC.(CRM-SYLMAR) UNIFY CRT-D; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3231-40Q
Medical Device Problem Code Inappropriate or Unexpected Reset (2959)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/04/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the implantable cardioverter defibrillator (icd) exhibited backup vvi due to a hardware reset.Programming changes were performed to resolve the issue.There were no patient consequences.
 
Event or Problem Description
New information indicates that the icd was explanted and replaced.
 
Additional Manufacturer Narrative
The reported event of backup operation was confirmed by reviewing the data received from technical services.The cause of the backup event was a power-on-reset (por).However, the cause of the por could not be determined due to lack of data.Functional testing was performed.Telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were tested and found to be normal.The backup operation could not be reproduced.The cause of the reported event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY CRT-D
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15655823
Report Number2017865-2022-42688
Device Sequence Number2428611
Product Code NIK
UDI-Device Identifier05414734504393
UDI-Public05414734504393
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P030054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2012
Device Model NumberCD3231-40Q
Device Catalogue NumberCD3231-40Q
Device Lot Number3401115
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/04/2022
Supplement Date Received by Manufacturer10/12/2022
01/23/2023
Initial Report FDA Received Date10/21/2022
Supplement Report FDA Received Date11/09/2022
01/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/24/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
1258T/86 PKG/STER.; 2088/TC 52 US TENDRILACTIVE.; 7120Q/58 DURATA.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexMale
-
-