| Catalog Number |
0375532000 |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/06/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of the device broke during the procedure.The tip was retrieved.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of the device broke during the procedure.The tip was retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned for investigation and the reported failure mode was confirmed.Alleged failure: the inner blade got stuck and stopped spinning.They found the tip was torn and broken off.The probable root cause/s could be the blade coming in contact with a hard object causing damage to the teeth and hitting the housing edges.Excessive pressure, such as bending or prying, may cause the arthroscopic shaver blades to bend/break.The failure mode will be monitored for future reoccurrence.
|
| |
|