The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents/complaints from this lot.The investigation was unable to determine a conclusive cause for the reported material rupture.Factors that may contribute to a material rupture include, but are not limited to, inadequate balloon integrity, interaction with challenging anatomy, interaction with accessory devices, and/or inflating above the rated burst pressure.In this case, based on the information provided and because the device was not returned for analysis, it is unknown what may have caused the reported material rupture.There is no indication of a product quality issue with respect to manufacture, design, or labeling.D9, h3 - device not available for evaluation.
|