SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Catalog Number C1034J
Device Problem Component Missing (2306)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/17/2019
Event Type  malfunction  
Manufacturer Narrative
This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).(udi) is unknown.No product information has been provided to date.510k is unknown.No information has been provided to date.A review of the device history records shows there were no observations recorded during manufacture to suggest an issue of this nature would occur with this lot of product.A product sample was received for evaluation.Visual and functional testing were performed.One sample was received in unused conditions, within the original packaging.The sample was visually inspected with a 12 inches distance under normal conditions of illumination.During the visual inspection, it was found that sample did not have the breathing filter.The root cause was related to manufacturing process.As a preventive action manufacturing personnel were notified as an awareness of the failure mode reported by the customer.
 
Event Description
It was reported that before opening the package, the customer noticed there was not a filter attached to the product.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX GENERAL ANESTHESIA CIRCUITS
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15656290
MDR Text Key306918266
Report Number3012307300-2022-25911
Device Sequence Number1
Product Code BTO
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/12/2022
Device Catalogue NumberC1034J
Device Lot Number3772406
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/29/2022
Initial Date FDA Received10/22/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/14/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-