| Model Number |
L-270 |
| Medical Device Problem Codes |
Break (1069); Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/20/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation.Visual and functional testing were performed.No physical damage was observed.The visual investigation found that the sample connector is broken.It is unlikely that the damage was caused in the manufacturing process; the most probable cause is that the damage was caused after the product left manufacturing facilities.The root cause of the reported issue was unable to be determined.Actions were taken to mitigate the reported issue: no corrective actions are required since it's unlikely that needle was bent during the manufacturing process.
|
| |
|
Event or Problem Description
|
|
It was reported that the device had a breakage and was leaking.No patient injury was reported.
|
| |
|