Model Number 470057-08
Device Problem Optical Distortion (3000)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/22/2022
Event Type  malfunction  
Event Description
It was reported that during a da vinci-assisted incisional hernia surgical procedure, the 30 degree endoscope had double vision.There was no reported injury.Intuitive surgical, inc.(isi) made multiple follow up attempts to obtain additional information.However, no further details have been received as of the date of this report.
 
Manufacturer Narrative
The 30 degree endoscope has not been returned to intuitive surgical, inc.(isi) for evaluation.If additional information is received, a follow-up mdr will be submitted.The 30 degree endoscope part# 470057-08/ lot/serial# (b)(4) was last used on (b)(6) 2022 on system (b)(4) with 1943 use remaining.No image or procedure video was provided for review.Based on the information available at this time, this complaint is being classified as a reportable event due to the following conclusion: it was reported that the endoscope displayed double vision.The issue was resolved through the use of a backup endoscope.Poor vision and/or camera control could result in tissue damage.At this time, the root cause of the failure is unknown.While there was no harm or injury to the patient, the reported failure mode could likely cause or contribute to an adverse event if it were to recur.
 
Event Description
Refer to h10/h11 for follow-up information.
 
Manufacturer Narrative
Additional information found in: d9, g3, g6, h2, and h3 evaluation information found in: h6 and h10 the 30-degree endoscope has been returned and evaluated by the failure analysis team.Failure analysis could not replicate nor reproduce the reported issue.The image was good in both eyes.Sapphire distal window was found to be damaged/cracked.Camera adapter and scope had friction issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NONE
Type of Device
ENDOSCOPE PLUS
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key15656382
MDR Text Key307097366
Report Number2955842-2022-14776
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier00886874116555
UDI-Public(01)00886874116555
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191736
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 09/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/22/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number470057-08
Device Catalogue Number470057
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Manufacturer Received11/14/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/09/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
DA VINCI INSTRUMENTS AND ACCESSORIES
-
-