Model Number 3006
Medical Device Problem Code Energy Output Problem (1431)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the device during an unspecified procedure did not seem to cauterize as efficiently to stop bleeding.The device at the time of the event is being used with a third party generator ( valleylab ft10).The procedure was completed with a completely different set of equipment, with similar device and with no further issues.No delay in the procedure was reported.There was no patient harm, no user injury reported due to the event.
 
Additional Manufacturer Narrative
The subject device has not yet been received for evaluation.The cause of the reported issue is unknown at this time.Supplemental report(s) will be submitted should any relevant new information is available.Investigation is ongoing.This report will be supplemented accordingly following investigation.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 8 months since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, a definitive root cause of the reported event could not be identified.However, it's likely the cause is related to an insufficient generator output level.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUT/FCP,5MM/33CM DES III (5/PK)
Common Device Name
ELECTROSURGICAL HANDPIECE/ELECTRODE SINGLE-USE
Manufacturer (Section D)
GYRUS ACMI, INC
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer Contact
todd brill
9600 louisiana avenue north
brooklyn park, MN 55445
5082077661
MDR Report Key15656413
Report Number3005975494-2022-00056
Device Sequence Number14961624
Product Code GEI
UDI-Device Identifier00821925035898
UDI-Public00821925035898
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K023492
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 11/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3006
Device Lot NumberFR220528
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/28/2022
Supplement Date Received by Manufacturer10/27/2022
Initial Report FDA Received Date10/23/2022
Supplement Report FDA Received Date11/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/24/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
VALLEYLAB FT10
Patient SexUnknown
-
-