ALCON RESEARCH, LLC - HUNTINGTON CLAREON VIVITY EXTENDED VISION HYDROPHOBIC IOL; EXTENDED DEPTH OF FOCUS INTRAOCULAR LENS Back to Search Results
Model Number CNWET0
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/21/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A sample device was not returned for analysis.Product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
An ophthalmologist reported that during a cataract extraction with an intraocular lens (iol) implant procedure, small iridescent, almost metallic like flecks on and within the lens itself was found.Additional information was received stating issue with complaint product was noticed upon insertion into eye.It was reported that there was no patient harm.
 
Additional Manufacturer Narrative
Additional information was provided in h.3, h.6 and h.10.File indicates the use of a qualified cartridge and viscoelastic.The file also indicates the use of a non-qualified handpiece.The product investigation could not identify a root cause.Without the sample returned for evaluation we are unable to determine if the reported deposits are present.The manufacturer internal reference number is:(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON VIVITY EXTENDED VISION HYDROPHOBIC IOL
Common Device Name
EXTENDED DEPTH OF FOCUS INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15656497
Report Number1119421-2022-02259
Device Sequence Number11338491
Product Code POE
UDI-Device Identifier00380652405331
UDI-Public00380652405331
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/05/2026
Device Model NumberCNWET0
Device Catalogue NumberCNWET0.245
Device Lot Number15369112
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/28/2022
Supplement Date Received by Manufacturer11/28/2022
Initial Report FDA Received Date10/23/2022
Supplement Report FDA Received Date12/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/07/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH II IOL, INJECTOR; MONARCH III IOL, CARTRIDGE D; PROVISC
Patient Age69 YR
Patient SexFemale
-
-