Catalog Number NPFS02000
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/28/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, during set up for a navio assisted surgery, when trying to power on the navio surgical system, they kept receiving an error code of ups back up error code not working properly.Rep tried leaving it on plugged in charged but still wouldn¿t start.Everything was plugged in.Rep then tried bypassing the ups and plugging it in the wall but still didn¿t work.The procedure was completed with manual instrumentations without delays.The patient was not harmed.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
H3, h6: the navio surgical system ups,(b)(6), intended for use in treatment was returned for evaluation.A relationship between the reported event and the device was established.Nothing was identified visually that contributed to the reported problem.A functional evaluation was performed.The reported problem was confirmed.The ups was plugged into ac power and will not power on.The ups has experienced a catastrophic failure.It is an oem device and further evaluation is not possible.The most likely cause of this event is an internal electronic failure of the ups rendering it completely nonfunctional.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints found one similar event.A historical escalation event review was completed.A review of prior escalation actions found no actions applicable to the scope of this case.The navio surgical technique guide for knee arthroplasty provides guidelines for recovering to a fully manual procedure in the ¿recovery procedure guidelines¿.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still adequate.Per complaint details from the us, it was reported that during set up for a navio assisted surgery, when trying to power on the navio surgical system, they kept receiving an error code of ups back up error code not working properly.The procedure was completed with manual instrumentation without delays.The patient was not harmed.Although no further containment or corrective action is recommended or required at this time, the failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO SURGICAL SYSTEM US
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15656516
Report Number3010266064-2022-00634
Device Sequence Number2303489
Product Code OLO
UDI-Device Identifier00885556628416
UDI-Public885556628416
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberNPFS02000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/28/2022
Supplement Date Received by Manufacturer12/01/2022
Initial Report FDA Received Date10/23/2022
Supplement Report FDA Received Date12/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;