| Model Number |
DRIVER SHAFT, T-15, LONG |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
During returned device evaluation, a reportable malfunction was discovered.Complaint confirmed.Upon visual inspection it was noted that the drive geometry at the distal tip of the returned ar-9545-t15-03 had both twisted and broken.No fragments were returned for inspection.It was not indicated if a torque-indicating adapter had been used with the device.A probable cause is attributed to over-torquing/over-engaging the driver within the screw head.
|
| |
|
Event or Problem Description
|
|
It was reported that during an arthroscopy the screw attachments got twisted.There was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.During returned device evaluation, a reportable malfunction was discovered.
|
| |
|