Model Number DRIVER SHAFT, T-15, MEDIUM
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
During returned device evaluation, a reportable malfunction was discovered.Complaint confirmed.Upon visual inspection it was noted that the drive geometry at the distal tip of the returned ar-9545-t15-02 had both twisted and broken.No fragments were returned for inspection.It was not indicated if a torque-indicating adapter had been used with the device.A probable cause is attributed to over-torquing/over-engaging the driver within the screw head.
 
Event or Problem Description
It was reported on (b)(6) 2022 by a sales representative via sems that an ar-9545-t15-02 and ar-9545-t15-03 driver shaft are twisted.This was discovered during an unspecified time with no patient harm.During returned device evaluation, a reportable malfunction was discovered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRIVER SHAFT, T-15, MEDIUM
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key15657386
Report Number1220246-2022-05649
Device Sequence Number12356792
Product Code LXH
UDI-Device Identifier00888867225190
UDI-Public00888867225190
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 10/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDRIVER SHAFT, T-15, MEDIUM
Device Catalogue NumberAR-9545-T15-02
Device Lot Number1392140
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/06/2022
Initial Report FDA Received Date10/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/30/2021
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-