| Model Number |
DRIVER SHAFT, T-15, MEDIUM |
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
08/03/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported on (b)(6) 2022 by a sales representative via sems that an ar-9545-t15-02 driver shaft and ar-9545-t15-03 driver shaft is stripped.This was discovered during a case with no patient harm.During returned device evaluation, a reportable malfunction was discovered.
|
| |
|
Additional Manufacturer Narrative
|
|
During returned device evaluation, a reportable malfunction was discovered.Complaint confirmed.Upon visual inspection it was noted that the drive geometry at the distal tip of the returned ar-9545-t15-02 had both twisted and broken.No fragments were returned for inspection.It was not indicated if a torque-indicating adapter had been used with the device.A probable cause is attributed to over-torquing/over-engaging the driver within the screw head.
|
| |
|