ARTHREX, INC. SUTR ANCH PEEK S-TAK KNOTLESS 3X12.7MM; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Model Number SUTR ANCH PEEK S-TAK KNOTLESS 3X12.7MM
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that during a shoulder stabilization 360° surgery the suture of the device snapped at knotless splice during final tensioning, before desired tension was achieved.Shuttling was performed gently using two cannulas.The damaged suture was retrieved from the patient.According to the surgeon no harm for patient, operator or third party occurred.The surgery was finished successfully.Update (b)(6) 2022.It was confirmed that the surgery was finished successfully with the same device.It was not necessary to switch the surgical technique or do a second surgery.
 
Additional Manufacturer Narrative
Complaint confirmed.The returned brc-02-01l was used to confirm the presence of broken suture but the cause of this event could not be determined from the information provided or the returned device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUTR ANCH PEEK S-TAK KNOTLESS 3X12.7MM
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key15657691
Report Number1220246-2022-05664
Device Sequence Number6387077
Product Code MAI
UDI-Device Identifier00888867024229
UDI-Public00888867024229
Combination Product (Y/N)N
Initial Reporter CountryEI
PMA/510(K) Number
K180594
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 12/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model NumberSUTR ANCH PEEK S-TAK KNOTLESS 3X12.7MM
Device Catalogue NumberAR-1938PS
Device Lot Number14565084
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/06/2022
Supplement Date Received by Manufacturer10/06/2022
Initial Report FDA Received Date10/24/2022
Supplement Report FDA Received Date12/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-