• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR TREK CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR TREK CORONARY DILATATION CATHETER Back to Search Results
Model Number 1012280-15
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the procedure was to treat a lesion in an unspecified artery.The 5.0x15mm trek rx balloon dilatation catheter (bdc) was used for post-dilatation, however, the balloon was unable to be removed after deflation.The trek rx bdc was removed together with the other devices as a unit.Another 5.0x15mm trek rx bdc was used and was unable to be removed again after deflation.The trek rx bdc was removed together with the guide wire as a unit.There was no adverse patient effect and there was no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device is expected to be returned for evaluation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.The additional trek rx device is being filed under a separate medwatch report number.
 
Additional Manufacturer Narrative
Visual and dimensional inspections were performed on the returned device.The reported difficulty removing the guide wire from the balloon catheter could not be confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported complaint.However, factors that may contribute to the difficult removing the guide wire causing resistance between the devices may include, but not limited to, manufacturing damage, inner diameter of guide wire lumen, condition of the guide wire, or damage to the distal shaft of the catheter.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15657726
Report Number2024168-2022-10949
Device Sequence Number17758165
Product Code LOX
UDI-Device Identifier08717648138782
UDI-Public08717648138782
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K103110
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2024
Device Model Number1012280-15
Device Catalogue Number1012280-15
Device Lot Number10226G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/04/2022
Supplement Date Received by Manufacturer11/07/2022
Initial Report FDA Received Date10/24/2022
Supplement Report FDA Received Date11/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-