A search for non-conformance's associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was stated to be available for return to the manufacturer for evaluation, but has not yet been returned.The event as described could not be confirmed.If the device is received at a later date, an investigation will be performed and a supplemental mdr will be submitted.The instructions for use (ifu) identifies premature coil detachment as a potential complication associated with the use of the device.
|
|
Please note this follow up report is also being submitted to include correction from fei based to cfn based.Fei was selected in error on our initial report.A visual inspection of the device captured in this file could not be performed as a physical device was not returned for evaluation, nor were any images of the device provided in place of a device return.Procedure and medical imaging were not provided for this investigation.Without the return and physical evaluation of the device, the investigation cannot definitively determine if a condition existed that would have caused or contributed to the reported event.Without imaging, the investigation cannot verify the event occurred as described, nor could the investigation definitively determine the cause of the reported event.Based on a review of the devices risk documentation, the reported event did not indicate there were any potential or new manufacturing, design, quality, or other systemic issues, or non-conformances.The complaint code is monitored through the trending process; corrective action is determined, as needed, through this process.Investigations of historic complaint files with similar complaint category coding are recorded in the complaint handling system, without the ability to perform and analysis of the device, this investigation cannot identify with certainty any potential root causes.A search for non-conformances associated with this part and lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.There is 1 other complaint from the last 2 years recorded in the complaint handling system with this batch number at the time of this investigation.At the time of this investigation, no systemic issues have been identified for this batch number that would have caused or contributed to the reported event.The instructions for use identifies premature coil detachment as a potential complication associated with the use of the device.
|