• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET 360 NANO 2MM X 4CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET 360 NANO 2MM X 4CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Model Number M0035442040
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/23/2022
Type of Reportable Event Malfunction
Event or Problem Description
During coil embolization of superior mesenteric artery for gastrointestinal bleeding, the subject coil had resistance while advancing past the introducer sheath.The physician successfully delivered the subject coil into the lesion.When detachment was attempted, the subject coil failed to detach.When the physician attempted to remove the subject coil along with the microcatheter, the subject coil detached prematurely into the intended vessel.The procedure was completed successfully with no clinical consequences to the patient.
 
Additional Manufacturer Narrative
The device is not available to the manufacturer.
 
Event or Problem Description
During coil embolization of superior mesenteric artery for gastrointestinal bleeding, the subject coil had resistance while advancing past the introducer sheath.The physician successfully delivered the subject coil into the lesion.When detachment was attempted, the subject coil failed to detach.When the physician attempted to remove the subject coil along with the microcatheter, the subject coil detached prematurely into the intended vessel.The procedure was completed successfully with no clinical consequences to the patient.
 
Additional Manufacturer Narrative
H4 manufacturing date ¿ added.D4 expiration date ¿ added.The subject device was not returned; therefore, visual and functional testing as well as physical analysis cannot be performed.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.Due to the automated manufacturing execution system there are controls in the manufacturing process to ensure the product met specifications upon release.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.Additional information provided by the customer indicated that the device was confirmed to be in good condition during preparation/prior to use on the patient and was prepared as per the dfu, continuous flush was set up and maintained throughout the clinical procedure, the patient's anatomy was very tortuous, the tapered distal end of the introducer sheath was firmly seated in the hub of the microcatheter, the rhv was adequately tightened while the introducer sheath was in the hub (not over/under-tightened) and was opened enough to allow passage of the device through without damage, no torque was given where advancing the delivery wire, and the coil was advanced slowly and gently without applying excessive force.Since the detached coil was placed in the treatment area, no medical intervention was required to remove it.While there are a number of potential causes for the reported issue, because review of available information failed to identify a definitive cause and the device was not returned for analysis, an assignable cause of undeterminable was assigned to the as reported events main coil prematurely detached/separated during use and coil introducer sheath friction.H3 other text : the device is not available to the manufacturer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET 360 NANO 2MM X 4CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key15657911
Report Number3008881809-2022-00522
Device Sequence Number5715486
Product Code HCG
UDI-Device Identifier07613252600299
UDI-Public07613252600299
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/20/2025
Device Model NumberM0035442040
Device Catalogue NumberM0035442040
Device Lot Number23496461
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/26/2022
Supplement Date Received by Manufacturer11/18/2022
Initial Report FDA Received Date10/24/2022
Supplement Report FDA Received Date11/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARNELIAN MARVEL MICROCATHETER (TOKAI MEDICAL); CARNELIAN MARVEL MICROCATHETER (TOKAI MEDICAL); INZONE DETACHMENT SYSTEM (STRYKER); INZONE DETACHMENT SYSTEM (STRYKER)