MAUDE Adverse Event Report: ST PAUL MEDFUSION 3500 SYRINGE INFUSION PUMP; PUMP, INFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL MEDFUSION 3500 SYRINGE INFUSION PUMP; PUMP, INFUSION Back to Search Results
Model Number 3500-0600-82
Medical Device Problem Code Defective Alarm (1014)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/10/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr was generated under protocol b10009704 as a result of warning letter cms# 617147.Manufacturing device history record review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.No causes or potential causes of the customer's reported problem were found during the review of service and repair records.Visual inspection did not notice any external damage.During start up testing, the reported problem was duplicated.The root cause of the reported issue was found to be the normal wear and drift of the force sensor.The force sensor was out of specification.With no pressure applied to the force sensor the reading was -1.01 and count was at zero.Recalibrated the force sensor and that cleared up the problem.Replaced force sensor as a preventative measure.A corrective and preventative action has been opened to address the reported issue.
 
Event or Problem Description
It was reported that there was a force sensor error.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDFUSION 3500 SYRINGE INFUSION PUMP
Common Device Name
PUMP, INFUSION
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15657999
Report Number3012307300-2022-26045
Device Sequence Number15061880
Product Code FRN
UDI-Device Identifier10610586043017
UDI-Public10610586043017
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K890120
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3500-0600-82
Device Catalogue Number3500-0600-82
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/20/2019
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/28/2022
Initial Report FDA Received Date10/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/10/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-