| Model Number |
900076 |
| Medical Device Problem Codes |
Break (1069); Defective Component (2292)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Trigger lever gets stuck.Fs log# 99264.1216677-2022-00285 wallach ultra freeze 900076 (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Coopersurgical, inc.Is currently investigating the reported condition.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation findings: the complaint product was manufactured at csi on 03-nov-2017 and sold on 28-dec-2017.Device history record was reviewed, and no non-conformities related to the complaint condition were noted.A review of the two-year complaint history showed similar reported complaint conditions.The complaint unit was returned on a repair on 10/oct/2022.Visual examination of the complaint product revealed no physical damage.Complaint product was functionally evaluated and found not to function properly.Root cause: while no definitive root cause could be reliably determined, the potential cause may be from normal wear and tear.Corrective actions: the complaint unit was repaired and returned to the customer.The relief device was replaced.The unit was tested and found acceptable.Coopersurgical will continue to trend this complaint condition.
|
| |
|
Event or Problem Description
|
|
Company representative confirmed "relief device not sealing to allow unit to build pressure.".
|
| |
|