|
|
| Model Number |
SDC140H |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
09/02/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Reason for device explant and/or reoperation: deflation.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a female patient underwent a breast surgery with 600cc mentor artoura plus, smooth, high profile saline tissue expanders.Post-operatively, the patient experienced bilateral deflation, which was confirmed during a physical exam.As a result, the patient underwent removal and replacement with unspecified non-mentor breast implants on (b)(6) 2022.Several follow-ups have been conducted, and no additional information has been received.If more information becomes available, mentor will submit a supplemental report accordingly.Refer to manufacturing report number 1645337-2022-12762 for the contralateral event.
|
| |
|
Additional Manufacturer Narrative
|
|
On november 14, 2022, mentor completed a visual inspection, leak testing, and microscopic examination of the returned device.Visual analysis of the returned sample revealed that no damage or anomalies were observed on the tissue expander.Leak testing was performed, in accordance with mentor procedures, and a tear was noted on the union between shell and anterior patch.Microscopic examination was performed and the cause of the deflation could not be identified.As the authorization form for examination was not received the product evaluation lab couldn't identify a conclusion due to the specific nature of this deflation.The evaluation was limited to non-destructive testing.Although no conclusion could be reach on the cause of the reported event, the instructions for use contain the following caution: causes of deflation of the tissue expander include but are not limited to the following events: crease fold failure, intraoperative or postoperative trauma, excessive stresses or manipulations as may occur during normal, daily routines including customary and purposeful trauma such as that which can occur during vigorous exercise, athletics, etc.Manufacturer's reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On november 08, 2022, the mentor failure analysis lab received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer's reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|