| Model Number |
301-07-10 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/09/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The instrument could have been broken during the surgery.However, the breakage has been noticed at the recovery of the instrumentation-set by our logistics dept., after the cleaning realized by the sterilization team of the hospital.No information has been received from the hospital regarding this issue and xnov ip didn¿t receive any request for complementary information regarding this issue as well.
|
| |
|
Additional Manufacturer Narrative
|
|
Pending investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: the broken instrument reported may have been the result of a stress riser, secondary to years of surgical use or unauthorized disassembly of subcomponents for cleaning, which led to crack initiation, propagation, and ultimate failure of the black radel sphere subcomponent.
|
| |
|