• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. PRM STM INSRTR EXTRCT; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. PRM STM INSRTR EXTRCT; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 301-07-10
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pending investigation.
 
Event or Problem Description
The instrument could have been broken during the surgery.However, the breakage has been noticed at the recovery of the instrumentation-set by our logistics dept., after the cleaning realized by the sterilization team of the hospital.No information has been received from the hospital regarding this issue and xnov ip didn¿t receive any request for complementary information regarding this issue as well.
 
Additional Manufacturer Narrative
H3: the broken instrument reported may have been the result of unauthorized disassembly of subcomponents for cleaning and not properly re-assembling the subcomponents after cleaning, which led to the missing radel sphere and threads subcomponents.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRM STM INSRTR EXTRCT
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66th ct
gainesville, FL 32653
3523771140
MDR Report Key15660493
Report Number1038671-2022-01348
Device Sequence Number5868121
Product Code LXH
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number301-07-10
Device Catalogue Number301-07-10
Device Lot Number145103005
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/04/2022
Supplement Date Received by Manufacturer02/13/2023
07/22/2026
Initial Report FDA Received Date10/24/2022
Supplement Report FDA Received Date02/28/2023
07/27/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-