• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US EXPRESSEW III SUTURE PASSER W/O HOOK; SUTURE/NEEDLE PASSER, REUSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US EXPRESSEW III SUTURE PASSER W/O HOOK; SUTURE/NEEDLE PASSER, REUSABLE Back to Search Results
Model Number 214140
Device Problems Device-Device Incompatibility (2919); Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/25/2022
Event Type  malfunction  
Event Description
It was reported by sales rep that during a shoulder arthroscopy procedure on (b)(6)2022 , it was observed that the jaw on the expressew iii suture passer w/o hook device would not close.During in-house engineering evaluation, it was observed that the jaw on the device was loose that it could be moved up and down with the fingertips.Another like device was used to complete the procedure with a 10 minute delay.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: udi: (b)(4).The expiration date is currently unavailable.Reporter is a j&j sales representative.Investigation summary the complaint device was received and evaluated.Upon visual inspection, the device does not show any structural anomalies.The jaw is loose, it can be moved up and down with the fingertips.Due to the condition of the jaw, the functional test cannot be performed.As a potential cause cannot be associated to manufacturing, therefore a manufacturing record evaluation is not required.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection result, this complaint can be confirmed.The possible root cause for the loose jaw condition can be attributed to procedural variables, such handling of the device or product interaction during procedure, a bigger portion of tissue may have been grabbed and forced the jaw to close, also, since this device is reusable, the continuous use and sterilization process can cause metal fatigue leading to a loose jaw, however, this cannot be conclusively determined.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESSEW III SUTURE PASSER W/O HOOK
Type of Device
SUTURE/NEEDLE PASSER, REUSABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key15660757
MDR Text Key306924293
Report Number1221934-2022-03264
Device Sequence Number1
Product Code LXH
UDI-Device Identifier10886705020218
UDI-Public10886705020218
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number214140
Device Catalogue Number214140
Device Lot Number48684
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/27/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/24/2022
Initial Date FDA Received10/24/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/27/2018
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-