SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
|
Back to Search Results |
|
Catalog Number C49101320J |
Device Problem
Material Split, Cut or Torn (4008)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/01/2018 |
Event Type
malfunction
|
Event Description
|
It was reported that immediately upon the use of the product, the product got torn.No patient injury was reported.
|
|
Manufacturer Narrative
|
This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).Udi is unknown.No product information has been provided to date.510k is unknown.No information has been provided to date.No lot number was provided; therefore, device history record review could not be completed.A product sample was received for evaluation.Visual and functional testing were performed.One sample was received without its original packaging inside a ziploc bag.The sample was visually inspected with a 12 inches distance under normal conditions of illumination.During the visual inspection, a tear was found in the tuber.The root cause was unable to be determined, most likely damaged after leaving the manufacturing facility.As a preventive action production personnel was notified by quality engineer as awareness of the defect reported by the customer.
|
|