| Model Number |
SRL-320 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/27/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture.The device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformance's noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: rupture: observed opening device in the posterior side assessed as surgical damage.Additional observations: observed creases on the device.Observed wear abrasion on the device.Stress marks observed on the device.White particles observed in the surface of the shell.No further actions are required as the device was implanted.
|
| |
|
Event or Problem Description
|
|
The device has been explanted.
|
| |
|