| Model Number |
E801 |
| Medical Device Problem Codes |
Low Test Results (2458); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The initial reporter complained of discrepant results for 1 patient sample tested for elecsys ferritin (ferritin) on a cobas e801 module.The initial result was 0.500 ng/ml with a data flag.The repeat result from the same instrument was 58.4 ng/ml.The repeat result from another instrument was 60.5 ng/ml.No questionable results were reported outside of the laboratory.The ferritin reagent lot number and expiration date were not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Section b3, date of event was updated.Section d4, serial number was updated.The ferritin reagent lot expiration date was jun-2023.Calibration and qc were acceptable.There was no indication of a general reagent issue.Instrument performance testing from 21-nov-2022 passed.Based on the information available, the cause of the event could not be determined.The investigation did not identify a product problem.
|
| |
|
Additional Manufacturer Narrative
|
|
The ferritin reagent lot was 602731.
|
| |
|