MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COBAS E 801 ANALYTICAL UNIT; IMMUNOCHEMISTRY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COBAS E 801 ANALYTICAL UNIT; IMMUNOCHEMISTRY ANALYZER Back to Search Results
Model Number E801
Medical Device Problem Codes Low Test Results (2458); Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/25/2022
Type of Reportable Event Malfunction
Event or Problem Description
The initial reporter complained of discrepant results for 1 patient sample tested for elecsys ferritin (ferritin) on a cobas e801 module.The initial result was 0.500 ng/ml with a data flag.The repeat result from the same instrument was 58.4 ng/ml.The repeat result from another instrument was 60.5 ng/ml.No questionable results were reported outside of the laboratory.The ferritin reagent lot number and expiration date were not provided.
 
Additional Manufacturer Narrative
Section b3, date of event was updated.Section d4, serial number was updated.The ferritin reagent lot expiration date was jun-2023.Calibration and qc were acceptable.There was no indication of a general reagent issue.Instrument performance testing from 21-nov-2022 passed.Based on the information available, the cause of the event could not be determined.The investigation did not identify a product problem.
 
Additional Manufacturer Narrative
The ferritin reagent lot was 602731.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS E 801 ANALYTICAL UNIT
Common Device Name
IMMUNOCHEMISTRY ANALYZER
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
HITACHI HIGH TECH CORP.
882 ichige hitachinaka
na
ibaraki 312-8 504
JA   312-8504
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15662747
Report Number1823260-2022-03324
Device Sequence Number7141758
Product Code JJE
UDI-Device Identifier04015630939688
UDI-Public04015630939688
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K060373
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberE801
Device Catalogue Number08454345001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/01/2022
Supplement Date Received by Manufacturer10/28/2022
12/12/2022
Initial Report FDA Received Date10/24/2022
Supplement Report FDA Received Date11/14/2022
12/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1