| Catalog Number |
03337154001 |
| Medical Device Problem Codes |
High Test Results (2457); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable roche omni electrode ph results from 3 patient samples on a cobas b 221 roche omni s6 system.For sample 1, the ph result was 7.45.The result on an avl analyzer was 7.33.For sample 2, the ph result was 7.39.The result on an avl analyzer was 7.28.For sample 3, the ph result was 7.43.The result on an avl analyzer was 7.30.The initial results were reported outside of the laboratory.The repeat results from the avl analyzer were deemed correct.The electrode serial number and expiration date were requested but not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
No further data was provided.The investigation did not identify a product problem.The root cause of the event could not be determined.
|
| |
|