Model Number FX-410615
Medical Device Problem Codes Device Appears to Trigger Rejection (1524); Material Rupture (1546)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 08/12/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Healthcare professional reported "a right side capsular contracture baker grade iv with "pain", "fold reported with ultrasound on (b)(6) 2022" and "firmness." device remains implanted.
 
Additional Manufacturer Narrative
(b)(6).A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iv and crease folding of implant.
 
Event or Problem Description
Device explanted.
 
Event or Problem Description
Healthcare professional additionally reported right side "grossly ruptured".
 
Additional Manufacturer Narrative
Device evaluation: based on the device analysis grid, the assessments of the complaint are: capsular contracture: unable to observe as it is a medical event that is not related to the device.Crease/folding of implant: crease flat is observed on the device.Rupture: observed, broken on radius side assessed as unidentified (tear) opening (shell thickness within specification).Additional observations: red particles on the shell and gel are observed.Crease flat is observed.Wear abrasion is observed.No further actions are required as the device was implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key15663117
Report Number9617229-2022-18581
Device Sequence Number6470423
Product Code FTR
UDI-Device Identifier10888628003026
UDI-Public10888628003026
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
P040046
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/05/2021
Device Model NumberFX-410615
Device Catalogue NumberFX-410615
Device Lot Number3003366
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/30/2022
Supplement Date Received by Manufacturer10/26/2022
11/21/2022
Initial Report FDA Received Date10/24/2022
Supplement Report FDA Received Date11/21/2022
12/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/05/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
Patient SexFemale
-
-