• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD MAXPLUS¿ PRESSURE RATED EXTENSION SET WITH NEEDLELESS CONNECTOR AND Y-SITE; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD MAXPLUS¿ PRESSURE RATED EXTENSION SET WITH NEEDLELESS CONNECTOR AND Y-SITE; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number MPX5302-C
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that 2 bd maxplus¿ pressure rated extension sets with needleless connector and y-site had clogged tubing during use.The following information was provided by the initial reporter: "fluid does not flow through tubing.Has happened at least two times with different staff with the same lot.".
 
Additional Manufacturer Narrative
Device evaluated by mfr: a device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes, d10: returned to manufacturer on: 06-oct-2022.H6: investigation summary: one sample model mpx5302-c was returned for investigation.The set was examined for defects and abnormalities.No defects or abnormalities were observed.A 10 ml bd syringe filled with water was attached to the set and the set was attempted to be primed.The set was unable to be primed.Upon closer visual inspection under a microscope, excess solvent was seen near the female luer.A device history record review for model mpx5302-c lot number 22069498 was performed.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.A trend for this occlusion issue has been identified for this product line.A capa (corrective action preventative action) has been initiated, and a team has been assembled in order to investigate the issue.
 
Event or Problem Description
It was reported that 2 bd maxplus¿ pressure rated extension sets with needleless connector and y-site had clogged tubing during use.The following information was provided by the initial reporter: "fluid does not flow through tubing.Has happened at least two times with different staff with the same lot.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MAXPLUS¿ PRESSURE RATED EXTENSION SET WITH NEEDLELESS CONNECTOR AND Y-SITE
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15663327
Report Number9616066-2022-01599
Device Sequence Number17766981
Product Code FPA
UDI-Device Identifier10885403237232
UDI-Public(01)10885403237232
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K051499
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue NumberMPX5302-C
Device Lot Number22069498
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/06/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/28/2022
Supplement Date Received by Manufacturer02/22/2023
Initial Report FDA Received Date10/24/2022
Supplement Report FDA Received Date02/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-