| Catalog Number |
MPX5302-C |
| Medical Device Problem Code |
Complete Blockage (1094)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that 2 bd maxplus¿ pressure rated extension sets with needleless connector and y-site had clogged tubing during use.The following information was provided by the initial reporter: "fluid does not flow through tubing.Has happened at least two times with different staff with the same lot.".
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: a device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields were updated due to additional information: d10: device available for eval yes, d10: returned to manufacturer on: 06-oct-2022.H6: investigation summary: one sample model mpx5302-c was returned for investigation.The set was examined for defects and abnormalities.No defects or abnormalities were observed.A 10 ml bd syringe filled with water was attached to the set and the set was attempted to be primed.The set was unable to be primed.Upon closer visual inspection under a microscope, excess solvent was seen near the female luer.A device history record review for model mpx5302-c lot number 22069498 was performed.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.A trend for this occlusion issue has been identified for this product line.A capa (corrective action preventative action) has been initiated, and a team has been assembled in order to investigate the issue.
|
| |
|
Event or Problem Description
|
|
It was reported that 2 bd maxplus¿ pressure rated extension sets with needleless connector and y-site had clogged tubing during use.The following information was provided by the initial reporter: "fluid does not flow through tubing.Has happened at least two times with different staff with the same lot.".
|
| |
|