Model Number 37612
Device Problems Delayed Charge Time (2586); Charging Problem (2892); Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/20/2022
Event Type  malfunction  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient was having difficulty getting all 8 coupling bars to fill in resulting in it taking longer to fully charge the implantable neurostimulator (ins).It used to take 10-12 minutes and now it takes 30 minutes.This morning, they initially had 0 coupling bars filled for the first couple minutes but were eventually able to get 6 coupling bars to fill in after repositioning the antenna.The patient added that they also rotated antenna dial.The patient confirmed they were able to maintain 6 to 8 coupling bars after that.The patient mentioned that they hold the recharger antenna in place with their hand.During the call, the coupling fluctuated between 6 and 8 bars.It was reviewed 6 to 8 bars is good coupling/charge efficiency.The patient examined the recharger antenna but no issue was found.The patient had previously reported similar coupling issue that resulted in the recharger being replaced.The agent reviewed that since the coupling issue occurred on two separate rechargers, there did not appear to be an issue with the recharger.The patient's next appointment with their hcp will be in (b)(6) 2023.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received from the patient stating that the patient had trouble getting all 8 bars to light up.The patient was able to resolve this by putting in a new charging unit in.The patient reports that the new charging unit works great now.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key15663676
MDR Text Key302593425
Report Number3004209178-2022-13892
Device Sequence Number1
Product Code MHY
UDI-Device Identifier00643169529762
UDI-Public00643169529762
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 11/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/24/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/28/2017
Device Model Number37612
Device Catalogue Number37612
Was Device Available for Evaluation? No
Date Manufacturer Received11/11/2022
Date Device Manufactured01/10/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age70 YR
Patient SexMale
Patient Weight86 KG
-
-