CUE HEALTH INC. CUE COVID-19 TEST FOR HOME AND OVER THE COUNTER (OTC) USE; REAGENTS, 2019-NOVEL CORONAVIRUS NUCLEIC ACID Back to Search Results
Model Number C2020
Device Problem False Positive Result (1227)
Patient Problems Sneezing (2251); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 02/06/2022
Event Type  malfunction  
Event Description
Customer reporting on behalf of relative (daughter).Customer reported an individual received a false positive result on (b)(6) 2022 using cue covid-19 test for home and over the counter (otc) use (cartridge sn (b)(4) followed by a negative result (cartridge sn (b)(4) on cue reader sn (b)(4) with cartridge lot 20857e on the same day.Customer reported the individual's sars-cov-2 pcr test, by color health, was negative on (b)(6) 2022.The non-vaccinated individual was in close contact with exposure on (b)(6) 2022.The individual presented with symptoms (runny nose and occasional sneezing).Customer confirmed cartridges were stored in the appropriate temperature range per package insert.
 
Manufacturer Narrative
Device evaluated by manufacturer: cartridges used by the consumer were not sent back for evaluation, therefore, the devices could not be evaluated.The customer provided information regarding the suspected false positive tests and a member of the technical support group was able to perform data analysis.The complaint history was reviewed and there was one similar complaint against involved lot.The lot history record (lhr) was reviewed for the lot number in the complaint, and it passed release specifications.A technical support representative reviewed customer data and performed data analysis.All data values were normal and met acceptable criteria.Root cause was undetermined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUE COVID-19 TEST FOR HOME AND OVER THE COUNTER (OTC) USE
Type of Device
REAGENTS, 2019-NOVEL CORONAVIRUS NUCLEIC ACID
Manufacturer (Section D)
CUE HEALTH INC.
4980 carroll canyon rd.
suite 100
san diego CA 92121
Manufacturer Contact
roderick castillo
4980 carroll canyon rd.
suite 100
san diego, CA 92121
8332838378
MDR Report Key15664116
MDR Text Key306708239
Report Number3016758165-2022-00180
Device Sequence Number1
Product Code QJR
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/24/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date09/14/2022
Device Model NumberC2020
Device Lot Number20857E
Was Device Available for Evaluation? No
Distributor Facility Aware Date02/08/2022
Date Manufacturer Received02/08/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/14/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-